Eoin O'Brien a Blood Pressure Unit, Beaumont Hospital,
Dublin 9, Ireland, b Centre Hospitalier Universitaire
Vaudois, Division D'Hypertension, Departement de medecine interne,
1011 Lausanne, Switzerland, c Universita degli Studi di Milano-Biocca, Istituto Scientifico
Ospedale San Luca, IRCCS, Istituto Auxologico Italiano, Via Spagnoletto
3, 20149 Milan, Italy, d Katholieke Universiteit Leuven,
Hypertensie en Cardiovasculaire Revalidatie Eenheid, Inwendige
Geneeskunde-Cardiologie, UZ Gasthuisberg, 3000 Leuven, Belgium, e Division of Cardiology,
Sunnybrook and Women's College Health Sciences Centre, 2075 Bayview
Avenue, Toronto, ON, Canada M4N 3M5
Correspondence to: E O'Brien eobrien{at}iol.ie
There is a large market for blood
pressure measuring devices not only in clinical medicine but also among
the public where the demand for self measurement of blood pressure is
growing rapidly. For consumers, whether medical or lay, accuracy should
be of prime importance when selecting a device to measure blood
pressure. However, most devices have not been evaluated for accuracy
independently using the two most widely used protocols: the British
Hypertension Society (BHS) protocol and the standard set by the US
Association for the Advancement of Medical Instrumentation
(AAMI).
1 2
The Working Group on Blood Pressure Monitoring
of the European Society of Hypertension has decided to review blood
pressure measuring devices regularly to guide purchasers.3
For this first report devices for which there is published evidence of
independent validation using these protocols have been surveyed.
Because most blood pressure devices have not been independently
validated, only a fraction of the many devices available have been
surveyed. Devices that have been validated recently for which results
have not yet been published were not included, but this shortcoming
should be addressed in future.
Validation standards
Table 1.
Summary points
Two manual sphygmomanometers have been validated, one is
recommended
Five devices for clinical use in hospitals have been validated, two are
recommended
23 devices for self measurement of blood pressure have been validated,
five are recommended
24 devices for ambulatory measurement of blood pressure have been
validated, 16 are recommended
Validations and recommendations will be updated on the
BMJ 's website
![]()
Methods
Top
Methods
Results
Discussion
References
In 1987, the American Association for the Advancement of
Medical Instrumentation published a standard for sphygmomanometers
which included a protocol for evaluating the accuracy of
devices.4 In 1990 a protocol was devised by the British
Hypertension Society.5 Both protocols have since been revised.
1 2
Since the two protocols can be reconciled the joint criteria are applied in most validation studies.6
The criteria for fulfilling the BHS protocol are that devices must achieve at least grade B (where A denotes greatest agreement with mercury standard and D denotes least agreement) for systolic and for
diastolic pressures (table 1)1; the criteria for
fulfilling the AAMI protocol are that the test device must not differ
from the mercury standard by a mean difference >5 mm Hg or a
standard deviation >8 mm
Hg.2
Criteria for recommendation
The following criteria have been used to designate devices
according to accuracy. A device is classed as recommended if it fulfils
the AAMI criteria for both systolic and diastolic pressures (denoted as
passed) and received a grade of A or B under the BHS protocol for both
systolic and diastolic blood pressures. A device is not recommended if
it fails the AAMI criteria for either systolic or diastolic pressure
and achieves a grade of C or D for either systolic or diastolic
pressure under the BHS protocol. A questionable recommendation is made
when there is doubt about the strength of evidence. This may occur when
a device fulfils the criteria of one protocol but not the other,
and it may be best not to recommend the device for clinical use until a
confirmatory study has been performed; when the validation results were
presented only in an abstract without sufficient detail to appraise the
methodology, and it may be best to withhold an opinion until the full
results have been published or at least provided to a potential
purchaser by the manufacturer; when the conditions of the protocols
have not been fully adhered to (listed as a protocol violation in our
scheme); or when a device fulfils the AAMI criteria for intra-arterial
validation (the BHS protocol does not advocate validation using direct
intra-arterial measurement1), but it may be best to await
a validation against indirect blood pressure measurement before
recommending the device for general clinical use.
Identification of devices
This review was based on two previous surveys (which should be
consulted for early validation studies that are not reproduced in this
review),
7 8
and computerised search programs were used to
identify validation studies in the literature published up to December
1999. Blood pressure measuring devices were divided into two broad
categories: manual sphygmomanometers, which include mercury and
anaeroid devices; and automated sphygmomanometers, which include
devices for clinical use in hospitals, for self measurement of blood
pressure, for ambulatory blood pressure measurement, and for measuring
blood pressure in community settings. (Information on manufacturers
appears on the BMJ 's website.)
| |
Results |
|---|
|
|
|---|
Manual sphygmomanometers
These devices are listed in table 2
.9-12 One model of the many mercury
sphygmomanometers available, the PyMah (PyMah Corporation, Flemington,
NJ), has been validated according to both protocols and is
recommended.9 Because mercury sphygmomanometers are
generally of a simple basic design with standard components, it is
probably reasonable to assume that most would be of similar accuracy.
The standard anaeroid sphygmomanometer has only been formally validated
according to the calibration procedure of the BHS
protocol,12 and the results support reservations about
anaeroid devices because they are susceptible to becoming inaccurate
without this being apparent to the
user.13
|
Automated sphygmomanometers
Devices for use in hospitals
Devices for clinical use in hospitals are listed in
table 3.14-18
|
Devices for self measurement of blood pressure
There are a large number of automated devices for self measurement
of blood pressure, virtually all of which use the oscillometric
technique. Formerly these devices used automated inflation and
deflation of a cuff applied to the upper arm over the brachial artery.
Recently this technique has been used to measure blood pressure over
the radial artery at the wrist, but since these devices become
inaccurate if the arm is not kept at heart level during measurement,
the working group is reluctant to recommend them regardless of
accuracy.19 Devices that measure blood pressure by
occluding a digital artery in the finger are also available, but
because the problem of limb position is even more critical and there is
the additional problem of peripheral vasoconstriction affecting
accuracy, this technique is no longer recommended. These devices have
not been considered in this review.19
|
|
Devices for ambulatory measurement of blood pressure
There are two techniques for measuring ambulatory blood
pressure: the commonly used method of intermittent measurement over 24 hours and the developing method of continuous waveform analysis.
Devices dependent on intermittent
measurement are listed in table 6.33-68 Many of these
devices have been validated for use in specific groups, such as elderly
people and pregnant women, and in differing circumstances, such as
during exercise and in different postures. Validating devices for use in ambulatory conditions presents many methodological difficulties, and
some evidence suggests that inaccuracies found during static conditions
may be amplified in ambulatory
conditions.69
|
The Portapres (TNO, Amsterdam), a portable recorder for
24 hour ambulatory monitoring, can provide beat to beat monitoring which gives waveform measurements similar to intra-arterial recordings. However, the technique is subject to various inaccuracies, which the
use of correction factors and digital filters in the latest model may
remove; this model is awaiting formal validation.
70 71
Devices for measuring blood pressure in a community setting
There is little information available on the accuracy of automated
devices that are installed in public retail areas
such as pharmacies,
supermarkets, health clinics, and companies in a variety of
industries
which permit the public to measure blood pressure without
charge in an unsupervised, crowded setting with high ambient noise.
Evaluations of one such device, the Vita Stat 90550 (Spacelabs Medical,
Redmond, WA), have had conflicting results.
72 73
| |
Discussion |
|---|
|
|
|---|
Manufacturers of blood pressure measuring devices use innovative technology to provide an array of systems that can analyse, store, and display features of a haemodynamic variable in ways that would have been beyond the dreams of the pioneers of the technique. Although the selection of a blood pressure measuring device may be influenced by many factors, a fundamental requirement must be that it gives accurate measurements; too often accuracy has been sacrificed for technological ingenuity.
The evidence from validation studies is accumulating, and devices are being scrutinised more critically; this has been the case with ambulatory devices used in specific populations, such as in children, elderly people, and pregnant women. However, the evidence is not always clear cut. There may have been protocol violations; the data published may have been inadequate, such as sometimes occurs when only abstracts have been published; and there may be disagreement between validation studies of the same device. None the less certain recommendations can be made to assist potential purchasers.
In interpreting the recommendations made by this survey the following factors should be considered. A device fulfilling the AAMI criteria and graded A or B for both systolic and diastolic pressure under the BHS protocol has been recommended on grounds of accuracy without equivocation; one that fails the AAMI protocol for either systolic or diastolic pressure and has a grade of C or D for either systolic or diastolic pressure under the BHS protocol cannot be recommended on the grounds of accuracy. Devices are given a questionable recommendation if there is an element of doubt in interpreting the results of a validation study. One circumstance that a purchaser should also consider, but for which we cannot make a recommendation, is the occasional conflict that arises when a device fulfils the criteria of the protocols when validated at one centre but not another. When this occurs the details of the methodology may need to be scrutinised to determine if differences in the selection of participants, for example, might explain the conflict; it may be best to await the results of a confirmatory study before deciding whether the device is accurate.
Only a fraction of the devices available worldwide have been independently validated. This is especially true of devices used for self measurement . In 1994, Ng and Small surveyed 423 automated devices, of which 161 were designed for self measurement.74 Since then the number of devices available for self measurement has increased greatly but comparatively few have been validated. The situation is even worse for automated devices designed for use in specialised areas of hospitals, such as operating theatres and intensive care units, where accuracy should be a priority. Only five of the hundreds of devices available have been validated using the two protocols, of which only two met the criteria for recommendation in this review. 8 75 The much used anaeroid sphygmomanometer has only recently been independently evaluated.12 However, because these devices become inaccurate with use without this inaccuracy being apparent to the user, it is also necessary to validate them after they have been used for some time.13
A serious dilemma is how to influence manufacturers to modify devices that have been shown to be inaccurate. The Dinamap Portable Monitor model 8100 (Critikon, Tampa, FL) is an example of this: despite a number of reports of inaccuracy 18 76 it is one of the most popular automated devices used in clinical practice and hypertension research. It seems that purchasers and users of expensive devices for blood pressure measurement in specialised hospital areas are prepared to accept the word of manufacturers with regard to their accuracy and performance and to ignore warnings from the scientific literature as to their shortcomings.
Again, the situation is worse for self measurement devices. Despite the poor accuracy record of devices that measure blood pressure at the wrist and the serious misgivings voiced by clinicians about these devices,19 their popularity is growing. In Germany, for example, 1.2 million self measurement devices are sold annually.30
The European Union and international organisations of specialists in
hypertension have unanimously recommended that all devices for
measuring blood pressure should be independently
validated.
19 77-82
The reality is, however, that most
devices are not validated independently. This may be partly due to the
expense of conducting validation studies using complex
protocols.
1 2
Recently, the European Society of
Hypertension Working Party on Blood Pressure Monitoring agreed
proposals to simplify the BHS protocol without compromising the
integrity of the procedures, and an international protocol for
validation is being drafted.83 This will help
manufacturers to market devices worldwide, expedite validation
procedures, reduce the expense of performing studies, and permit more
centres to undertake validation procedures; all of which would enable
manufacturers to have all devices validated independently before they
are marketed. Early publication of validation studies might further
encourage manufacturers to have their devices evaluated, and the
readiness of Blood Pressure Monitoring to act as a
repository of peer reviewed studies is welcomed. The internet might
provide a means of continuously updating information on blood pressure
measuring devices.
| |
Footnotes |
|---|
Funding: None.
Competing interests: EO'B is director of the blood pressure unit at Beaumont Hospital, which has been contracted by manufacturers from time to time to conduct validation studies of blood pressure measurement devices; the results of these studies have been published. EO'B has advised AccuSphyg, which is developing a non-mercury automated device for use in hospitals, and he holds a minority financial interest in the company.
Additional information about
manufacturers and members of the working group appear on the BMJ's
website
Members of the working group are Roland Asmar, Andrew Coats, Denis L Clement, Peter de Leeuw, Robert Fagard, Yutaka Imai, Jean-Michel Mallion, Giuseppe Mancia, Thomas Mengden, Martin Myers, Eoin O'Brien, Paul Padfield, Gianfranco Parati, Paolo Palatini, Thomas G Pickering, Josep Redon, Jan Staessen, Peter Trenkwalder, Gert van Montfrans, Paolo Verdecchia, Bernard Waeber, William White
| |
References |
|---|
|
|
|---|
| 1. | O'Brien E, Petrie J, Littler WA, de Swiet M, Padfield PL, Altman D, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11(suppl 2): S43-S63. |
| 2. | Association for the Advancement of Medical Instrumentation. American national standard. Electronic or automated sphygmomanometers. ANSI/AAMI SP 10-1992. Arlington, VA: AAMI, 1993:40. |
| 3. | O'Brien E. Formation of the Working Group on Blood Pressure Monitoring of the European Society of Hypertension. Blood Press Monit 1998; 3: 133-134. |
| 4. | Association for the Advancement of Medical Instrumentation. American national standard for electronic or automated sphygmomanometers. ANSI/AAMI SP10-1987. Arlington, VA: AAMI, 1987:25. |
| 5. | O'Brien E, Petrie J, Littler WA, Padfield PL, O'Malley K, Jamieson M, et al. British hypertension protocol: evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990; 8: 607-619[Medline]. |
| 6. | O'Brien E, Atkins N. A comparison of the BHS and AAMI protocols for validating blood pressure measuring devices: can the two be reconciled? J Hypertens 1994; 12: 1089-1094[Medline]. |
| 7. |
O'Brien E, Atkins N, Staessen J.
State of the market: a review of ambulatory blood pressure monitoring devices.
Hypertension
1995;
26:
835-842 |
| 8. | O'Brien E. Automated blood pressure measurement: state of the market in 1998 and the need for an international validation protocol for blood pressure measuring devices. Blood Press Monit 1998; 3: 205-211[Medline]. |
| 9. | O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of seven popular sphygmomanometers for home-measurement of blood pressure. J Hypertens 1990; 8: 621-634[CrossRef][Medline]. |
| 10. | O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of the Hawksley random-zero sphygmomanometer. Lancet 1990; 336: 1465-1468[CrossRef][Medline]. |
| 11. | Brown WCB, Kennedy S, Inglis GC, Murray LS, Lever AF. Mechanisms by which the Hawksley random zero sphygmomanometer underestimates blood pressure and produces a non-random distribution of RZ values. J Hum Hypertens 1997; 11: 75-93[CrossRef][Medline]. |
| 12. | Lones CR, Khanna M, Rushbrook J, Poston L, Shennan AH. Are aneroid devices suitable replacements for mercury sphygmomanometers? J Hum Hypertens 2000; 14: 843. |
| 13. | Burke MJ, Towers HM, O'Malley K, Fitzgerald D, O'Brien E. Sphygmomanometers in hospitals and family practice: problems and recommendations. BMJ 1982; 285: 469-471. |
| 14. | Anwar YA, Tendler BE, McCabe EJ, Mansoor GA, White WB. Evaluation of the Datascope Accutorr Plus according to the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1997; 2: 105-110[Medline]. |
| 15. | Alpert BS. Validation of CAS model 9010 automated blood pressure monitor: children/adult and neonatal studies. Blood Press Monit 1996; 1: 69-73[Medline]. |
| 16. | Bobrie G, Battaglia C. Assessment of Tensionic Mod EPS 112 semi-automatic device of blood pressure measurement (Manufactured by Spengler - France) [Abstract]. J Hypertens 1999; 17(suppl 3): S22-S23. |
| 17. | Zorn EA, Wilson MB, Angel JJ, Zanella J, Alpert B. Validation of an automated arterial tonometry monitor using Association for the Advancement of Medical Instrumentation standards. Blood Press Monit 1997; 2: 185-188[Medline]. |
| 18. | O'Brien E, Mee F, Atkins N, O'Malley K. Short report: accuracy of the Dinamap portable monitor, model 8100 determined by the British Hypertension Society protocol. J Hypertens 1993; 11: 761-763[CrossRef][Medline]. |
| 19. | O'Brien E, De Gaudemaria R, Bobrie G, Agabiti Rosei E, Vaisse B, the participants of the first international consensus conference on blood pressure self-measurement. Devices and validation. Blood Press Monit 2000; 5: 93-100[Medline]. |
| 20. | O'Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Press Monit 1996; 1: 55-61[Medline]. |
| 21. | Foster C, McKinlay JM, Cruickshank JM, Coats AJS. Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Hum Hypertens 1994; 8: 661-664[Medline]. |
| 22. | Walma EP, van Dooren C, van der Does E, Prins A, Mulder P, Hoes AW. Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. J Hum Hypertens 1995; 9: 169-174[Medline]. |
| 23. | Rithalia SVS, Edwards D. Comparison of oscillometric and intra-arterial blood pressure and pulse measurement. J Med Eng Technol 1994; 18: 179-181[Medline]. |
| 24. | Artigao M, Llavador J, Rubio M, Torres C, Lopez J, Sanchis C, et al. Evaluation of two devices for self-measurement of blood pressure according to the Britsih Hypertension Society protocol: the Omron M4 and MX2 [Abstract]. Am J Hypertens 1998; 16: S276. |
| 25. | Córdoba R, Fuertes MI, Alvarez A, Molina L, Solans R, Melero I. Validation of a self-monitoring blood pressure monitor: the Omron-HM 722C. Aten Primaria 1997; 20: 247-250[Medline]. |
| 26. | Bortolotto LA, Henry O, Hanon O, Sikias P, Mourad J-J, Girerd X. Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. Blood Press Monit 1999; 4: 21-25[Medline]. |
| 27. | Mufunda J, Sparks B, Chifamba J, Dakwa C, Matenga JA, Adams JM, et al. Comparison of the Omron HEM-713C automated blood pressure monitor with a standard auscultatory method using a mercury manometer. Cent Afr J Med 1996; 42: 230-232[Medline]. |
| 28. | Anwar YA, Giacco S, McCabe EJ, Tendler BE, White WB. Evaluation of the efficacy of the Omron HEM-737 Intellisense device for use on adults according to the recommendations of the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1998; 3: 261-266[Medline]. |
| 29. | Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and `range of uncertainity' of oscillometric blood pressure monitors around the upper arm and wrist. Blood Press Monit 1998; 3: 339-346[Medline]. |
| 30. | Eckert S, Gleichmann U, Zagorski O, Klapp A. Validation of the Omron R3 blood pressure self-measuring device through simultaneous comparative invasive measurements according to protocol 58130 of the German Institute for Validation. Blood Press Monit 1997; 2: 189-192[Medline]. |
| 31. | Rogers P, Burke V, Stroud P, Puddey IB. Comparison of oscillometric blood pressure measurements at the wrist with an upper-arm auscultatory mercury sphygmomanometer. Clin Exp Pharmacol Physiol 1999; 26: 477-481[Medline]. |
| 32. | Shennan AH, Rushbrook J, Power J, Wright J, Poston L. An accurate oscillometric wrist blood pressure monitor: validation of the Omron Rx [HEM-608]. J Hum Hypertens 1998; 12: 794. |
| 33. | Taylor R, Chidley K, Goodwin J, Broeders M, Kirby B. Accutracker II (version 30/23) ambulatory blood pressure monitor: clinical validationusing the British Hypertension Society and Association for the Advancement of Medical Instrumentation standards. J Hypertens 1993; 11: 1275-1282[Medline]. |
| 34. | O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the CH- Druck/Pressure ERKA ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens 1993; 11(suppl 2): S1-S7[Medline]. |
| 35. | Livi R, Teghini L, Cagnoni S, Scarpelli PT. Simultaneous and sequential same-arm measurements in the validation studies of automated blood pressure measuring devices. Am J Hypertens 1996; 9: 1228-1231[Medline]. |
| 36. | O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society Protocol. J Hypertens 1991; 9(suppl 5): S9-15. |
| 37. | Gosse P, Laforge A, Ansoborto P, Julien P, Lemetayer P, Clementy J. Clinical evaluation of the DIASYS integra blood pressure recorder [Abstract]. J Hypertens 1997; 15(suppl): 18S. |
| 38. | Kuwajima I, Nishinaga M, Kanamaru A. The accuracy and clinical performance of a new compact ambulatory blood pressure monitoring device, the ES-H531. Am J Hypertens 1998; 11: 1328-1333[Medline]. |
| 39. | Manning G, Vijan SG, Millar-Craig MW. Technical and clinical evaluation of the Medilog ABP non-invasive blood pressure monitor. J Ambulatory Monitoring 1994; 7: 255-264. |
| 40. | Barna I, Keszei A, Dunai A. Evaluation of Meditech ABPM-04 ambulatory blood pressure measuring device according to the British Hypertension Society protocol. Blood Press Monit 1998; 3: 363-368[Medline]. |
| 41. | Mee F, Atkins N, O'Brien E. Validation of the Nissei DS-240 ambulatory blood pressure measuring system as determined by the British Hypertension Society protocol. J Hum Hypertens 1994; 8: 295. |
| 42. | Germanó G, Muscolo M, Angotti S, Bravo S, Codispoti P, Federico L, et al. Evaluation of a new ambulatory blood pressure device. Am J Hypertens 1998; 11: 1486-1491[Medline]. |
| 43. | O'Brien E, Mee F, Atkins N, O'Malley K. Short report: accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society Protocol. J Hypertens 1991; 9: 567-568[Medline]. |
| 44. | O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Profilomat ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens 1993; 11(suppl 22): S9-15[Medline]. |
| 45. | Franx A, van der Post JAM, van Montfrans GA, Bruinse HW. Comparison of an auscultatory versus an oscillometric ambulatory blood pressure monitor in normotensive, hypertensive and preeclamptic pregnancy. Hypertens Pregnancy 1997; 16: 187-202. |
| 46. | Mee F, Atkins N, O'Brien E. Evaluation of the Profilomat II ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999; 3: 353-361. |
| 47. | White WB, Susser W, James G, Marra L, McCabe EJ, Pickering TG, et al. Multicentre assessment of the Quiet Trak ambulatory blood pressure recorder according to the 1992 AAMI guidelines. Am J Hypertens 1994; 7: 509-514[Medline]. |
| 48. | O'Shea JC, O'Neill A, O'Brien C, Murphy MB. Evaluation of the Tycos Quiet Trak ambulatory pressure recorder using British Hypertension Society (BHS) protocol [Abstract]. Am J Hypertens 1994; 7: 118A. |
| 49. | Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol 1999; 181: 1203-1210[CrossRef][Medline]. |
| 50. | Penny JA, Shennan AH, Rushbrook J, Halligan A, Taylor DJ, de Swiet M. Validation of the Welch Allyn QuietTrak ambulatory blood pressure monitor in pregnancy. Hypertens Pregnancy 1996; 15: 313-321. |
| 51. | Modesti PA, Costoli A, Cecioni I, Toccafondi S, Carnemolla A, Serneri GGN. Clinical evaluation of the QuietTrak blood pressure recorder according to the protocol of the British Hypertension Society. Blood Press Monit 1996; 1: 63-68[Medline]. |
| 52. | De Gaudemaris R, White A, Pascal I, Siche J-P, Baguet J-P, Mallion J-M. Evaluation du moniteur de pression artérielle en ambulatoire Save 33 modèle 2. Arch Mal Coeur Vaiss 1999; 92: 1133-1137[Medline]. |
| 53. | Mee F, Atkins N, O'Brien E. Evaluation of the Shiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999; 4: 35-43[Medline]. |
| 54. | O'Brien E, Atkins N, Mee F, O'Malley K. Evaluation of the SpaceLabs 90202 according to the AAMI standard and BHS criteria. J Hum Hypertens 1991; 5: 223-226[Medline]. |
| 55. | O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the SpaceLabs 90207 determined by to the British Hypertension Society Protocol. J Hypertens 1991; 9(suppl 5): S25-S31. |
| 56. | O'Brien E, Mee F, Atkins N, Halligan A, O'Malley K. Accuracy of the SpaceLabs 90207 ambulatory blood pressure measuring system in normotensive pregnant women determined by the British Hypertension Society Protocol. J Hypertens 1993; 11(suppl 5): S282-S283. |
| 57. | Shennan AH, Kissane J, de Swiet M. Validation of the Spacelabs 90207 ambulatory blood pressure monitoring in pregnancy. Br J Obstet Gynaecol 1993; 100: 904-908[Medline]. |
| 58. | Shennan A, Haligan A, Gupta M, Taylor D, de Swiet M. Oscillometric blood pressure measurements in severe preeclampsia: validation of the SpaceLabs 90207. Br J Obstet Gynaecol 1996; 103: 171-173[Medline]. |
| 59. | Belsha CW, Wells TG, Rice HB, Neaville WA, Berry PL. Accuracy of the SpaceLabs 90207 ambulatory blood pressure monitor in children and adolescents. Blood Press Monit 1996; 1: 127-133[Medline]. |
| 60. | Iqbal P, Fotherby MD, Potter JF. Validation of the SpaceLabs 90207 automatic non-invasive blood pressure monitor in elderly subjects. Blood Press Monit 1996; 1: 367-373[Medline]. |
| 61. | Peixoto AJ, Gray TA, Crowley ST. Validation of the SpaceLabs 90207 ambulatory blood pressure device for hemodialysis patients. Blood Press Monit 1999; 4: 217-221[Medline]. |
| 62. | Baumgart P, Kamp J. Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor. Blood Press Monit 1998; 3: 303-307[Medline]. |
| 63. | O'Brien E, Mee F, Atkins N, O'Malley K. Short report: accuracy of the Takeda TM-2420/TM-2020 determined by the British Hypertension Society protocol. J Hypertens 1991; 9: 571-572[CrossRef][Medline]. |
| 64. | White WB, Pickering TG, Morganroth J, James GD, McCabe EJ, Moucha O, et al. A multicenter evaluation of the A&D TM-2420 ambulatory blood pressure recorder. Am J Hypertens 1991; 4: 890-896[Medline]. |
| 65. | Palatini P, Penzo M, Canali C, Pessina AC. Validation of the accuracy of the A&D TM2420 Model 7 for ambulatory blood pressure monitoring and effect of microphone replacement on its performance. J Ambul Monit 1991; 4: 281-288. |
| 66. | Imai Y, Sasaki S, Minami N, Munakakata M, Hashimoto J, Sakuma H, et al. The accuracy and performance of the A&D TM 2421, a new ambulatory blood pressure monitoring device based on the cuff-oscillometric method and the Korotkov sound technique. Am J Hypertens 1992; 5: 719-726[Medline]. |
| 67. | O'Sullivan JJ, Derrick G, Griggs PE, Wren C. Validation of the Takeda 2421 ambulatory blood pressure monitor in children. J Med Eng Technol 1998; 22: 101-105[Medline]. |
| 68. | Palatini P, Frigo G, Bertolo O, Roman E, Da Corta R, Winnicki M. Validation of the A&D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to subjects' characteristics. Blood Press Monit 1998; 3: 255-260[Medline]. |
| 69. |
Omboni S, Parati G, Groppelli A, Ulian L, Mancia G.
Performance of the AM-5600 blood pressure monitor: comparison with ambulatory intra-arterial pressure.
J Appl Physiol
1997;
82:
698-703 |
| 70. |
Omboni G, Parati P, Castiglioni M, Di Rienzo BP, Imholz GJ, Langewouters KH, Wesseling G, Mancia G.
Estimation of blood pressure variability from 24 hour ambulatory finger blood pressure.
Hypertension
1998;
32:
52-58 |
| 71. | Gizdulich P, Imholz BPM, van den Meiracker AH, Parati G, Wesseling KH. Finapres tracking of systolic blood pressure and baroreflex sensitivity improved by waveform filtering. J Hypertens 1996; 14: 243-250[Medline]. |
| 72. | Nara AR. Performance review of a noninvasive blood pressure monitor. Med Electron 1996; 27: 63-67. |
| 73. | Myers MG, Magharious L, Boyle E, Lewis JE. Accuracy of the Vita-Stat device in measuring blood pressure. Can J Cardiol 1998; 14(suppl F): 148F. |
| 74. | Ng K-G, Small CF. Survey of automated noninvasive blood pressure monitors. J Clin Eng 1994; 19: 452-475[Medline]. |
| 75. | Ng K-G. Automated noninvasive blood pressure measurement. Intensive Care World 1995; 12: 89-102[Medline]. |
| 76. | O'Brien E, Atkins N. Accuracy of the Dinamap portable monitor, model 8100: a review of the evidence for accuracy. Blood Press Monit 1997; 2: 31-33[Medline]. |
| 77. | Pickering T, for the American Society of Hypertension Ad Hoc Panel. Recommendations for the use of home (self) and ambulatory blood pressure monitoring. Am J Hypertens 1996; 9: 1-11[CrossRef][Medline]. |
| 78. |
O'Brien E, Coats A, Owens P, Petrie J, Padheld P, Littler WA, et al.
Use and interpretation of ambulatory blood pressure monitoring: recommendations of the British Hypertension Society.
BMJ
2000;
320:
1128-1134 |
| 79. | Myers MG, Haynes RB, Rabkin SW. Canadian Hypertension Society guidelines for ambulatory blood pressure monitoring. Am J Hypertens 1999; 12: 1149-1157[CrossRef][Medline]. |
| 80. | European Committee for Standardization. European standard EN 1060-1 (British standard BSSEN 1060-1:1996). Specification for non-invasive sphygmomanometers. Part 1. General requirements. Brussels: European Committee for Standardization, 1995. |
| 81. | European Committee for Standardization. European standard EN 1060-2 (British standard BSSEN 1060-2:1996). Specification for non-invasive sphygmomanometers. Part 2. Supplementary requirements for mechanical sphygmomanometers. Brussels: European Committee for Standardization, 1995. |
| 82. | European Committee for Standardization. Non-invasive sphygmomanometers. Part 3. Supplementary requirements for electro-mechanical blood pressure measuring systems. British Standard BS EN 1060-3: 1997. European Standard EN 1060-3 1997. Brussels: European Committee for Standardization, 1997. |
| 83. | O'Brien E. Proposals for simplyfing the validation protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 2000; 5: 43-45[Medline]. |
(Accepted 17 November 2000)
Read all Rapid Responses