BMJ 2001;322:531-536 ( 3 March )

Clinical review

Blood pressure measuring devices: recommendations of the European Society of Hypertension

Eoin O'Brien, professor of cardiovascular medicine aBernard Waeber, professor bGianfranco Parati, associate professor cJan Staessen, lecturer dMartin G Myers, professor e on behalf of the European Society of Hypertension Working Group on Blood Pressure Monitoring.

a Blood Pressure Unit, Beaumont Hospital, Dublin 9, Ireland, b Centre Hospitalier Universitaire Vaudois, Division D'Hypertension, Departement de medecine interne, 1011 Lausanne, Switzerland, c Universita degli Studi di Milano-Biocca, Istituto Scientifico Ospedale San Luca, IRCCS, Istituto Auxologico Italiano, Via Spagnoletto 3, 20149 Milan, Italy, d Katholieke Universiteit Leuven, Hypertensie en Cardiovasculaire Revalidatie Eenheid, Inwendige Geneeskunde-Cardiologie, UZ Gasthuisberg, 3000 Leuven, Belgium, e Division of Cardiology, Sunnybrook and Women's College Health Sciences Centre, 2075 Bayview Avenue, Toronto, ON, Canada M4N 3M5

Correspondence to: E O'Brien eobrien{at}iol.ie

There is a large market for blood pressure measuring devices not only in clinical medicine but also among the public where the demand for self measurement of blood pressure is growing rapidly. For consumers, whether medical or lay, accuracy should be of prime importance when selecting a device to measure blood pressure. However, most devices have not been evaluated for accuracy independently using the two most widely used protocols: the British Hypertension Society (BHS) protocol and the standard set by the US Association for the Advancement of Medical Instrumentation (AAMI). 1 2 The Working Group on Blood Pressure Monitoring of the European Society of Hypertension has decided to review blood pressure measuring devices regularly to guide purchasers.3 For this first report devices for which there is published evidence of independent validation using these protocols have been surveyed. Because most blood pressure devices have not been independently validated, only a fraction of the many devices available have been surveyed. Devices that have been validated recently for which results have not yet been published were not included, but this shortcoming should be addressed in future.


Summary points


Two manual sphygmomanometers have been validated, one is recommended

Five devices for clinical use in hospitals have been validated, two are recommended

23 devices for self measurement of blood pressure have been validated, five are recommended

24 devices for ambulatory measurement of blood pressure have been validated, 16 are recommended

Validations and recommendations will be updated on the BMJ 's website




    Methods
Top
Methods
Results
Discussion
References

Validation standards
In 1987, the American Association for the Advancement of Medical Instrumentation published a standard for sphygmomanometers which included a protocol for evaluating the accuracy of devices.4 In 1990 a protocol was devised by the British Hypertension Society.5 Both protocols have since been revised. 1 2 Since the two protocols can be reconciled the joint criteria are applied in most validation studies.6 The criteria for fulfilling the BHS protocol are that devices must achieve at least grade B (where A denotes greatest agreement with mercury standard and D denotes least agreement) for systolic and for diastolic pressures (table 1)1; the criteria for fulfilling the AAMI protocol are that the test device must not differ from the mercury standard by a mean difference >5 mm Hg or a standard deviation >8 mm Hg.2


                              
View this table:
[in this window]
[in a new window]
 

Table 1. Grading criteria used by the British Society of Hypertension.1 Grades represent the cumulative percentage of readings falling within 5 mm Hg, 10 mm Hg, and 15 mm Hg of the mercury standard. All three percentages must be greater than or equal to the values shown for a specific grade to be awarded. Values are mm Hg

Criteria for recommendation
The following criteria have been used to designate devices according to accuracy. A device is classed as recommended if it fulfils the AAMI criteria for both systolic and diastolic pressures (denoted as passed) and received a grade of A or B under the BHS protocol for both systolic and diastolic blood pressures. A device is not recommended if it fails the AAMI criteria for either systolic or diastolic pressure and achieves a grade of C or D for either systolic or diastolic pressure under the BHS protocol. A questionable recommendation is made when there is doubt about the strength of evidence. This may occur when a device fulfils the criteria of one protocol but not the other, and it may be best not to recommend the device for clinical use until a confirmatory study has been performed; when the validation results were presented only in an abstract without sufficient detail to appraise the methodology, and it may be best to withhold an opinion until the full results have been published or at least provided to a potential purchaser by the manufacturer; when the conditions of the protocols have not been fully adhered to (listed as a protocol violation in our scheme); or when a device fulfils the AAMI criteria for intra-arterial validation (the BHS protocol does not advocate validation using direct intra-arterial measurement1), but it may be best to await a validation against indirect blood pressure measurement before recommending the device for general clinical use.

Identification of devices
This review was based on two previous surveys (which should be consulted for early validation studies that are not reproduced in this review), 7 8 and computerised search programs were used to identify validation studies in the literature published up to December 1999. Blood pressure measuring devices were divided into two broad categories: manual sphygmomanometers, which include mercury and anaeroid devices; and automated sphygmomanometers, which include devices for clinical use in hospitals, for self measurement of blood pressure, for ambulatory blood pressure measurement, and for measuring blood pressure in community settings. (Information on manufacturers appears on the BMJ 's website.)


    Results
Top
Methods
Results
Discussion
References

Manual sphygmomanometers
These devices are listed in table 2 .9-12 One model of the many mercury sphygmomanometers available, the PyMah (PyMah Corporation, Flemington, NJ), has been validated according to both protocols and is recommended.9 Because mercury sphygmomanometers are generally of a simple basic design with standard components, it is probably reasonable to assume that most would be of similar accuracy. The standard anaeroid sphygmomanometer has only been formally validated according to the calibration procedure of the BHS protocol,12 and the results support reservations about anaeroid devices because they are susceptible to becoming inaccurate without this being apparent to the user.13


                              
View this table:
[in this window]
[in a new window]
 

Table 2. Manual blood pressure measuring devices validated using the protocols of the Association for the Advancement of Medical Instrumentation and the British Hypertension Society

Automated sphygmomanometers

Devices for use in hospitals
Devices for clinical use in hospitals are listed in table 3.14-18


                              
View this table:
[in this window]
[in a new window]
 

Table 3. Automated blood pressure measuring devices for clinical use in hospitals validated using the protocols of the Association for the Advancement of Medical Instrumentation and the British Hypertension Society. Devices were validated in oscillometric mode unless otherwise indicated

Devices for self measurement of blood pressure
There are a large number of automated devices for self measurement of blood pressure, virtually all of which use the oscillometric technique. Formerly these devices used automated inflation and deflation of a cuff applied to the upper arm over the brachial artery. Recently this technique has been used to measure blood pressure over the radial artery at the wrist, but since these devices become inaccurate if the arm is not kept at heart level during measurement, the working group is reluctant to recommend them regardless of accuracy.19 Devices that measure blood pressure by occluding a digital artery in the finger are also available, but because the problem of limb position is even more critical and there is the additional problem of peripheral vasoconstriction affecting accuracy, this technique is no longer recommended. These devices have not been considered in this review.19

Automated devices for upper arm measurement are shown in table 4. 9 20-29 Automated devices for wrist measurement are listed in table 5.29-32 These have been validated against brachial arterial measurements.


                              
View this table:
[in this window]
[in a new window]
 

Table 4. Automated blood pressure measuring devices for self measurement at the upper arm validated using the protocols of the Association for the Advancement of Medical Instrumentation and the British Hypertension Society. For the first seven devices grading criteria had not been established although the British protocol was in use.13 Devices were validated in oscillometric mode unless otherwise indicated


                              
View this table:
[in this window]
[in a new window]
 

Table 5. Automated blood pressure measuring devices for self measurement at the wrist validated using the protocols of the Association for the Advancement of Medical Instrumentation and the British Hypertension Society

Devices for ambulatory measurement of blood pressure
There are two techniques for measuring ambulatory blood pressure: the commonly used method of intermittent measurement over 24 hours and the developing method of continuous waveform analysis.

Intermittent measurement---Devices dependent on intermittent measurement are listed in table 6.33-68 Many of these devices have been validated for use in specific groups, such as elderly people and pregnant women, and in differing circumstances, such as during exercise and in different postures. Validating devices for use in ambulatory conditions presents many methodological difficulties, and some evidence suggests that inaccuracies found during static conditions may be amplified in ambulatory conditions.69


                              
View this table:
[in this window]
[in a new window]
 

Table 6. Ambulatory blood pressure measuring devices validated using the protocols of the Association for the Advancement of Medical Instrumentation and the British Hypertension Society

Devices for continuous non-invasive blood pressure monitoring of a finger---The Portapres (TNO, Amsterdam), a portable recorder for 24 hour ambulatory monitoring, can provide beat to beat monitoring which gives waveform measurements similar to intra-arterial recordings. However, the technique is subject to various inaccuracies, which the use of correction factors and digital filters in the latest model may remove; this model is awaiting formal validation. 70 71

Devices for measuring blood pressure in a community setting
There is little information available on the accuracy of automated devices that are installed in public retail areas---such as pharmacies, supermarkets, health clinics, and companies in a variety of industries---which permit the public to measure blood pressure without charge in an unsupervised, crowded setting with high ambient noise. Evaluations of one such device, the Vita Stat 90550 (Spacelabs Medical, Redmond, WA), have had conflicting results. 72 73


    Discussion
Top
Methods
Results
Discussion
References

Manufacturers of blood pressure measuring devices use innovative technology to provide an array of systems that can analyse, store, and display features of a haemodynamic variable in ways that would have been beyond the dreams of the pioneers of the technique. Although the selection of a blood pressure measuring device may be influenced by many factors, a fundamental requirement must be that it gives accurate measurements; too often accuracy has been sacrificed for technological ingenuity.

The evidence from validation studies is accumulating, and devices are being scrutinised more critically; this has been the case with ambulatory devices used in specific populations, such as in children, elderly people, and pregnant women. However, the evidence is not always clear cut. There may have been protocol violations; the data published may have been inadequate, such as sometimes occurs when only abstracts have been published; and there may be disagreement between validation studies of the same device. None the less certain recommendations can be made to assist potential purchasers.

In interpreting the recommendations made by this survey the following factors should be considered. A device fulfilling the AAMI criteria and graded A or B for both systolic and diastolic pressure under the BHS protocol has been recommended on grounds of accuracy without equivocation; one that fails the AAMI protocol for either systolic or diastolic pressure and has a grade of C or D for either systolic or diastolic pressure under the BHS protocol cannot be recommended on the grounds of accuracy. Devices are given a questionable recommendation if there is an element of doubt in interpreting the results of a validation study. One circumstance that a purchaser should also consider, but for which we cannot make a recommendation, is the occasional conflict that arises when a device fulfils the criteria of the protocols when validated at one centre but not another. When this occurs the details of the methodology may need to be scrutinised to determine if differences in the selection of participants, for example, might explain the conflict; it may be best to await the results of a confirmatory study before deciding whether the device is accurate.

Only a fraction of the devices available worldwide have been independently validated. This is especially true of devices used for self measurement . In 1994, Ng and Small surveyed 423 automated devices, of which 161 were designed for self measurement.74 Since then the number of devices available for self measurement has increased greatly but comparatively few have been validated. The situation is even worse for automated devices designed for use in specialised areas of hospitals, such as operating theatres and intensive care units, where accuracy should be a priority. Only five of the hundreds of devices available have been validated using the two protocols, of which only two met the criteria for recommendation in this review. 8 75 The much used anaeroid sphygmomanometer has only recently been independently evaluated.12 However, because these devices become inaccurate with use without this inaccuracy being apparent to the user, it is also necessary to validate them after they have been used for some time.13

A serious dilemma is how to influence manufacturers to modify devices that have been shown to be inaccurate. The Dinamap Portable Monitor model 8100 (Critikon, Tampa, FL) is an example of this: despite a number of reports of inaccuracy 18 76 it is one of the most popular automated devices used in clinical practice and hypertension research. It seems that purchasers and users of expensive devices for blood pressure measurement in specialised hospital areas are prepared to accept the word of manufacturers with regard to their accuracy and performance and to ignore warnings from the scientific literature as to their shortcomings.

Again, the situation is worse for self measurement devices. Despite the poor accuracy record of devices that measure blood pressure at the wrist and the serious misgivings voiced by clinicians about these devices,19 their popularity is growing. In Germany, for example, 1.2 million self measurement devices are sold annually.30

The European Union and international organisations of specialists in hypertension have unanimously recommended that all devices for measuring blood pressure should be independently validated. 19 77-82 The reality is, however, that most devices are not validated independently. This may be partly due to the expense of conducting validation studies using complex protocols. 1 2 Recently, the European Society of Hypertension Working Party on Blood Pressure Monitoring agreed proposals to simplify the BHS protocol without compromising the integrity of the procedures, and an international protocol for validation is being drafted.83 This will help manufacturers to market devices worldwide, expedite validation procedures, reduce the expense of performing studies, and permit more centres to undertake validation procedures; all of which would enable manufacturers to have all devices validated independently before they are marketed. Early publication of validation studies might further encourage manufacturers to have their devices evaluated, and the readiness of Blood Pressure Monitoring to act as a repository of peer reviewed studies is welcomed. The internet might provide a means of continuously updating information on blood pressure measuring devices.

    Footnotes

   Funding: None.

Competing interests: EO'B is director of the blood pressure unit at Beaumont Hospital, which has been contracted by manufacturers from time to time to conduct validation studies of blood pressure measurement devices; the results of these studies have been published. EO'B has advised AccuSphyg, which is developing a non-mercury automated device for use in hospitals, and he holds a minority financial interest in the company.

Additional information about manufacturers and members of the working group appear on the BMJ's website

Members of the working group are Roland Asmar, Andrew Coats, Denis L Clement, Peter de Leeuw, Robert Fagard, Yutaka Imai, Jean-Michel Mallion, Giuseppe Mancia, Thomas Mengden, Martin Myers, Eoin O'Brien, Paul Padfield, Gianfranco Parati, Paolo Palatini, Thomas G Pickering, Josep Redon, Jan Staessen, Peter Trenkwalder, Gert van Montfrans, Paolo Verdecchia, Bernard Waeber, William White


    References
Top
Methods
Results
Discussion
References

1. O'Brien E, Petrie J, Littler WA, de Swiet M, Padfield PL, Altman D, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11(suppl 2): S43-S63.
2. Association for the Advancement of Medical Instrumentation. American national standard. Electronic or automated sphygmomanometers. ANSI/AAMI SP 10-1992. Arlington, VA: AAMI, 1993:40.
3. O'Brien E. Formation of the Working Group on Blood Pressure Monitoring of the European Society of Hypertension. Blood Press Monit 1998; 3: 133-134.
4. Association for the Advancement of Medical Instrumentation. American national standard for electronic or automated sphygmomanometers. ANSI/AAMI SP10-1987. Arlington, VA: AAMI, 1987:25.
5. O'Brien E, Petrie J, Littler WA, Padfield PL, O'Malley K, Jamieson M, et al. British hypertension protocol: evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990; 8: 607-619[Medline].
6. O'Brien E, Atkins N. A comparison of the BHS and AAMI protocols for validating blood pressure measuring devices: can the two be reconciled? J Hypertens 1994; 12: 1089-1094[Medline].
7. O'Brien E, Atkins N, Staessen J. State of the market: a review of ambulatory blood pressure monitoring devices. Hypertension 1995; 26: 835-842[Abstract/Free Full Text].
8. O'Brien E. Automated blood pressure measurement: state of the market in 1998 and the need for an international validation protocol for blood pressure measuring devices. Blood Press Monit 1998; 3: 205-211[Medline].
9. O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of seven popular sphygmomanometers for home-measurement of blood pressure. J Hypertens 1990; 8: 621-634[CrossRef][Medline].
10. O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of the Hawksley random-zero sphygmomanometer. Lancet 1990; 336: 1465-1468[CrossRef][Medline].
11. Brown WCB, Kennedy S, Inglis GC, Murray LS, Lever AF. Mechanisms by which the Hawksley random zero sphygmomanometer underestimates blood pressure and produces a non-random distribution of RZ values. J Hum Hypertens 1997; 11: 75-93[CrossRef][Medline].
12. Lones CR, Khanna M, Rushbrook J, Poston L, Shennan AH. Are aneroid devices suitable replacements for mercury sphygmomanometers? J Hum Hypertens 2000; 14: 843.
13. Burke MJ, Towers HM, O'Malley K, Fitzgerald D, O'Brien E. Sphygmomanometers in hospitals and family practice: problems and recommendations. BMJ 1982; 285: 469-471.
14. Anwar YA, Tendler BE, McCabe EJ, Mansoor GA, White WB. Evaluation of the Datascope Accutorr Plus according to the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1997; 2: 105-110[Medline].
15. Alpert BS. Validation of CAS model 9010 automated blood pressure monitor: children/adult and neonatal studies. Blood Press Monit 1996; 1: 69-73[Medline].
16. Bobrie G, Battaglia C. Assessment of Tensionic Mod EPS 112 semi-automatic device of blood pressure measurement (Manufactured by Spengler - France) [Abstract]. J Hypertens 1999; 17(suppl 3): S22-S23.
17. Zorn EA, Wilson MB, Angel JJ, Zanella J, Alpert B. Validation of an automated arterial tonometry monitor using Association for the Advancement of Medical Instrumentation standards. Blood Press Monit 1997; 2: 185-188[Medline].
18. O'Brien E, Mee F, Atkins N, O'Malley K. Short report: accuracy of the Dinamap portable monitor, model 8100 determined by the British Hypertension Society protocol. J Hypertens 1993; 11: 761-763[CrossRef][Medline].
19. O'Brien E, De Gaudemaria R, Bobrie G, Agabiti Rosei E, Vaisse B, the participants of the first international consensus conference on blood pressure self-measurement. Devices and validation. Blood Press Monit 2000; 5: 93-100[Medline].
20. O'Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Press Monit 1996; 1: 55-61[Medline].
21. Foster C, McKinlay JM, Cruickshank JM, Coats AJS. Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Hum Hypertens 1994; 8: 661-664[Medline].
22. Walma EP, van Dooren C, van der Does E, Prins A, Mulder P, Hoes AW. Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. J Hum Hypertens 1995; 9: 169-174[Medline].
23. Rithalia SVS, Edwards D. Comparison of oscillometric and intra-arterial blood pressure and pulse measurement. J Med Eng Technol 1994; 18: 179-181[Medline].
24. Artigao M, Llavador J, Rubio M, Torres C, Lopez J, Sanchis C, et al. Evaluation of two devices for self-measurement of blood pressure according to the Britsih Hypertension Society protocol: the Omron M4 and MX2 [Abstract]. Am J Hypertens 1998; 16: S276.
25. Córdoba R, Fuertes MI, Alvarez A, Molina L, Solans R, Melero I. Validation of a self-monitoring blood pressure monitor: the Omron-HM 722C. Aten Primaria 1997; 20: 247-250[Medline].
26. Bortolotto LA, Henry O, Hanon O, Sikias P, Mourad J-J, Girerd X. Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. Blood Press Monit 1999; 4: 21-25[Medline].
27. Mufunda J, Sparks B, Chifamba J, Dakwa C, Matenga JA, Adams JM, et al. Comparison of the Omron HEM-713C automated blood pressure monitor with a standard auscultatory method using a mercury manometer. Cent Afr J Med 1996; 42: 230-232[Medline].
28. Anwar YA, Giacco S, McCabe EJ, Tendler BE, White WB. Evaluation of the efficacy of the Omron HEM-737 Intellisense device for use on adults according to the recommendations of the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1998; 3: 261-266[Medline].
29. Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and `range of uncertainity' of oscillometric blood pressure monitors around the upper arm and wrist. Blood Press Monit 1998; 3: 339-346[Medline].
30. Eckert S, Gleichmann U, Zagorski O, Klapp A. Validation of the Omron R3 blood pressure self-measuring device through simultaneous comparative invasive measurements according to protocol 58130 of the German Institute for Validation. Blood Press Monit 1997; 2: 189-192[Medline].
31. Rogers P, Burke V, Stroud P, Puddey IB. Comparison of oscillometric blood pressure measurements at the wrist with an upper-arm auscultatory mercury sphygmomanometer. Clin Exp Pharmacol Physiol 1999; 26: 477-481[Medline].
32. Shennan AH, Rushbrook J, Power J, Wright J, Poston L. An accurate oscillometric wrist blood pressure monitor: validation of the Omron Rx [HEM-608]. J Hum Hypertens 1998; 12: 794.
33. Taylor R, Chidley K, Goodwin J, Broeders M, Kirby B. Accutracker II (version 30/23) ambulatory blood pressure monitor: clinical validationusing the British Hypertension Society and Association for the Advancement of Medical Instrumentation standards. J Hypertens 1993; 11: 1275-1282[Medline].
34. O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the CH- Druck/Pressure ERKA ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens 1993; 11(suppl 2): S1-S7[Medline].
35. Livi R, Teghini L, Cagnoni S, Scarpelli PT. Simultaneous and sequential same-arm measurements in the validation studies of automated blood pressure measuring devices. Am J Hypertens 1996; 9: 1228-1231[Medline].
36. O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society Protocol. J Hypertens 1991; 9(suppl 5): S9-15.
37. Gosse P, Laforge A, Ansoborto P, Julien P, Lemetayer P, Clementy J. Clinical evaluation of the DIASYS integra blood pressure recorder [Abstract]. J Hypertens 1997; 15(suppl): 18S.
38. Kuwajima I, Nishinaga M, Kanamaru A. The accuracy and clinical performance of a new compact ambulatory blood pressure monitoring device, the ES-H531. Am J Hypertens 1998; 11: 1328-1333[Medline].
39. Manning G, Vijan SG, Millar-Craig MW. Technical and clinical evaluation of the Medilog ABP non-invasive blood pressure monitor. J Ambulatory Monitoring 1994; 7: 255-264.
40. Barna I, Keszei A, Dunai A. Evaluation of Meditech ABPM-04 ambulatory blood pressure measuring device according to the British Hypertension Society protocol. Blood Press Monit 1998; 3: 363-368[Medline].
41. Mee F, Atkins N, O'Brien E. Validation of the Nissei DS-240 ambulatory blood pressure measuring system as determined by the British Hypertension Society protocol. J Hum Hypertens 1994; 8: 295.
42. Germanó G, Muscolo M, Angotti S, Bravo S, Codispoti P, Federico L, et al. Evaluation of a new ambulatory blood pressure device. Am J Hypertens 1998; 11: 1486-1491[Medline].
43. O'Brien E, Mee F, Atkins N, O'Malley K. Short report: accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society Protocol. J Hypertens 1991; 9: 567-568[Medline].
44. O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Profilomat ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens 1993; 11(suppl 22): S9-15[Medline].
45. Franx A, van der Post JAM, van Montfrans GA, Bruinse HW. Comparison of an auscultatory versus an oscillometric ambulatory blood pressure monitor in normotensive, hypertensive and preeclamptic pregnancy. Hypertens Pregnancy 1997; 16: 187-202.
46. Mee F, Atkins N, O'Brien E. Evaluation of the Profilomat II ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999; 3: 353-361.
47. White WB, Susser W, James G, Marra L, McCabe EJ, Pickering TG, et al. Multicentre assessment of the Quiet Trak ambulatory blood pressure recorder according to the 1992 AAMI guidelines. Am J Hypertens 1994; 7: 509-514[Medline].
48. O'Shea JC, O'Neill A, O'Brien C, Murphy MB. Evaluation of the Tycos Quiet Trak ambulatory pressure recorder using British Hypertension Society (BHS) protocol [Abstract]. Am J Hypertens 1994; 7: 118A.
49. Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol 1999; 181: 1203-1210[CrossRef][Medline].
50. Penny JA, Shennan AH, Rushbrook J, Halligan A, Taylor DJ, de Swiet M. Validation of the Welch Allyn QuietTrak ambulatory blood pressure monitor in pregnancy. Hypertens Pregnancy 1996; 15: 313-321.
51. Modesti PA, Costoli A, Cecioni I, Toccafondi S, Carnemolla A, Serneri GGN. Clinical evaluation of the QuietTrak blood pressure recorder according to the protocol of the British Hypertension Society. Blood Press Monit 1996; 1: 63-68[Medline].
52. De Gaudemaris R, White A, Pascal I, Siche J-P, Baguet J-P, Mallion J-M. Evaluation du moniteur de pression artérielle en ambulatoire Save 33 modèle 2. Arch Mal Coeur Vaiss 1999; 92: 1133-1137[Medline].
53. Mee F, Atkins N, O'Brien E. Evaluation of the Shiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999; 4: 35-43[Medline].
54. O'Brien E, Atkins N, Mee F, O'Malley K. Evaluation of the SpaceLabs 90202 according to the AAMI standard and BHS criteria. J Hum Hypertens 1991; 5: 223-226[Medline].
55. O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the SpaceLabs 90207 determined by to the British Hypertension Society Protocol. J Hypertens 1991; 9(suppl 5): S25-S31.
56. O'Brien E, Mee F, Atkins N, Halligan A, O'Malley K. Accuracy of the SpaceLabs 90207 ambulatory blood pressure measuring system in normotensive pregnant women determined by the British Hypertension Society Protocol. J Hypertens 1993; 11(suppl 5): S282-S283.
57. Shennan AH, Kissane J, de Swiet M. Validation of the Spacelabs 90207 ambulatory blood pressure monitoring in pregnancy. Br J Obstet Gynaecol 1993; 100: 904-908[Medline].
58. Shennan A, Haligan A, Gupta M, Taylor D, de Swiet M. Oscillometric blood pressure measurements in severe preeclampsia: validation of the SpaceLabs 90207. Br J Obstet Gynaecol 1996; 103: 171-173[Medline].
59. Belsha CW, Wells TG, Rice HB, Neaville WA, Berry PL. Accuracy of the SpaceLabs 90207 ambulatory blood pressure monitor in children and adolescents. Blood Press Monit 1996; 1: 127-133[Medline].
60. Iqbal P, Fotherby MD, Potter JF. Validation of the SpaceLabs 90207 automatic non-invasive blood pressure monitor in elderly subjects. Blood Press Monit 1996; 1: 367-373[Medline].
61. Peixoto AJ, Gray TA, Crowley ST. Validation of the SpaceLabs 90207 ambulatory blood pressure device for hemodialysis patients. Blood Press Monit 1999; 4: 217-221[Medline].
62. Baumgart P, Kamp J. Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor. Blood Press Monit 1998; 3: 303-307[Medline].
63. O'Brien E, Mee F, Atkins N, O'Malley K. Short report: accuracy of the Takeda TM-2420/TM-2020 determined by the British Hypertension Society protocol. J Hypertens 1991; 9: 571-572[CrossRef][Medline].
64. White WB, Pickering TG, Morganroth J, James GD, McCabe EJ, Moucha O, et al. A multicenter evaluation of the A&D TM-2420 ambulatory blood pressure recorder. Am J Hypertens 1991; 4: 890-896[Medline].
65. Palatini P, Penzo M, Canali C, Pessina AC. Validation of the accuracy of the A&D TM2420 Model 7 for ambulatory blood pressure monitoring and effect of microphone replacement on its performance. J Ambul Monit 1991; 4: 281-288.
66. Imai Y, Sasaki S, Minami N, Munakakata M, Hashimoto J, Sakuma H, et al. The accuracy and performance of the A&D TM 2421, a new ambulatory blood pressure monitoring device based on the cuff-oscillometric method and the Korotkov sound technique. Am J Hypertens 1992; 5: 719-726[Medline].
67. O'Sullivan JJ, Derrick G, Griggs PE, Wren C. Validation of the Takeda 2421 ambulatory blood pressure monitor in children. J Med Eng Technol 1998; 22: 101-105[Medline].
68. Palatini P, Frigo G, Bertolo O, Roman E, Da Corta R, Winnicki M. Validation of the A&D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to subjects' characteristics. Blood Press Monit 1998; 3: 255-260[Medline].
69. Omboni S, Parati G, Groppelli A, Ulian L, Mancia G. Performance of the AM-5600 blood pressure monitor: comparison with ambulatory intra-arterial pressure. J Appl Physiol 1997; 82: 698-703[Abstract/Free Full Text].
70. Omboni G, Parati P, Castiglioni M, Di Rienzo BP, Imholz GJ, Langewouters KH, Wesseling G, Mancia G. Estimation of blood pressure variability from 24 hour ambulatory finger blood pressure. Hypertension 1998; 32: 52-58[Abstract/Free Full Text].
71. Gizdulich P, Imholz BPM, van den Meiracker AH, Parati G, Wesseling KH. Finapres tracking of systolic blood pressure and baroreflex sensitivity improved by waveform filtering. J Hypertens 1996; 14: 243-250[Medline].
72. Nara AR. Performance review of a noninvasive blood pressure monitor. Med Electron 1996; 27: 63-67.
73. Myers MG, Magharious L, Boyle E, Lewis JE. Accuracy of the Vita-Stat device in measuring blood pressure. Can J Cardiol 1998; 14(suppl F): 148F.
74. Ng K-G, Small CF. Survey of automated noninvasive blood pressure monitors. J Clin Eng 1994; 19: 452-475[Medline].
75. Ng K-G. Automated noninvasive blood pressure measurement. Intensive Care World 1995; 12: 89-102[Medline].
76. O'Brien E, Atkins N. Accuracy of the Dinamap portable monitor, model 8100: a review of the evidence for accuracy. Blood Press Monit 1997; 2: 31-33[Medline].
77. Pickering T, for the American Society of Hypertension Ad Hoc Panel. Recommendations for the use of home (self) and ambulatory blood pressure monitoring. Am J Hypertens 1996; 9: 1-11[CrossRef][Medline].
78. O'Brien E, Coats A, Owens P, Petrie J, Padheld P, Littler WA, et al. Use and interpretation of ambulatory blood pressure monitoring: recommendations of the British Hypertension Society. BMJ 2000; 320: 1128-1134[Free Full Text].
79. Myers MG, Haynes RB, Rabkin SW. Canadian Hypertension Society guidelines for ambulatory blood pressure monitoring. Am J Hypertens 1999; 12: 1149-1157[CrossRef][Medline].
80. European Committee for Standardization. European standard EN 1060-1 (British standard BSSEN 1060-1:1996). Specification for non-invasive sphygmomanometers. Part 1. General requirements. Brussels: European Committee for Standardization, 1995.
81. European Committee for Standardization. European standard EN 1060-2 (British standard BSSEN 1060-2:1996). Specification for non-invasive sphygmomanometers. Part 2. Supplementary requirements for mechanical sphygmomanometers. Brussels: European Committee for Standardization, 1995.
82. European Committee for Standardization. Non-invasive sphygmomanometers. Part 3. Supplementary requirements for electro-mechanical blood pressure measuring systems. British Standard BS EN 1060-3: 1997. European Standard EN 1060-3 1997. Brussels: European Committee for Standardization, 1997.
83. O'Brien E. Proposals for simplyfing the validation protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 2000; 5: 43-45[Medline].

(Accepted 17 November 2000)


© BMJ 2001

Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

Relevant Article

Unvalidated blood pressure devices with small cuffs are being used in hospitals
Mark Thomas, Tamsin Radford, and Indranil Dasgupta
BMJ 2001 323: 398. [Extract] [Full Text]

This article has been cited by other articles:

  • Smith, R. D., Levy, P. J. (2008). Review: New techniques for assessment of vascular function. Therapeutic Advances in Cardiovascular Disease 2: 373-385 [Abstract]  
  • di Giuseppe, R., Di Castelnuovo, A., Centritto, F., Zito, F., De Curtis, A., Costanzo, S., Vohnout, B., Sieri, S., Krogh, V., Donati, M. B., de Gaetano, G., Iacoviello, L. (2008). Regular Consumption of Dark Chocolate Is Associated with Low Serum Concentrations of C-Reactive Protein in a Healthy Italian Population. J. Nutr. 138: 1939-1945 [Abstract] [Full text]  
  • Carroll, D., Phillips, A. C., Der, G. (2008). Body Mass Index, Abdominal Adiposity, Obesity, and Cardiovascular Reactions to Psychological Stress in a Large Community Sample. Psychosom. Med. 70: 653-660 [Abstract] [Full text]  
  • Pickering, T. G., Miller, N. H., Ogedegbe, G., Krakoff, L. R., Artinian, N. T., Goff, D. (2008). Call to Action on Use and Reimbursement for Home Blood Pressure Monitoring: A Joint Scientific Statement From the American Heart Association, American Society of Hypertension, and Preventive Cardiovascular Nurses Association. Hypertension 52: 10-29 [Abstract] [Full text]  
  • Green, B. B., Cook, A. J., Ralston, J. D., Fishman, P. A., Catz, S. L., Carlson, J., Carrell, D., Tyll, L., Larson, E. B., Thompson, R. S. (2008). Effectiveness of Home Blood Pressure Monitoring, Web Communication, and Pharmacist Care on Hypertension Control: A Randomized Controlled Trial. JAMA 299: 2857-2867 [Abstract] [Full text]  
  • Touze, E., Coste, J., Voicu, M., Kansao, J., Masmoudi, R., Doumenc, B., Durieux, P., Mas, J.-L. (2008). Importance of In-Hospital Initiation of Therapies and Therapeutic Inertia in Secondary Stroke Prevention: IMplementation of Prevention After a Cerebrovascular evenT (IMPACT) Study. Stroke 39: 1834-1843 [Abstract] [Full text]  
  • Chen, X., Wang, Y., Appel, L. J., Mi, J. (2008). Impacts of Measurement Protocols on Blood Pressure Tracking From Childhood Into Adulthood: A Metaregression Analysis. Hypertension 51: 642-649 [Abstract] [Full text]  
  • Minutolo, R., Borrelli, S., Scigliano, R., Bellizzi, V., Chiodini, P., Cianciaruso, B., Nappi, F., Zamboli, P., Conte, G., De Nicola, L. (2007). Prevalence and clinical correlates of white coat hypertension in chronic kidney disease. Nephrol Dial Transplant 22: 2217-2223 [Abstract] [Full text]  
  • Skov, K., Eiskjaer, H., Hansen, H. E., Madsen, J. K., Kvist, S., Mulvany, M. J. (2007). Treatment of Young Subjects at High Familial Risk of Future Hypertension With an Angiotensin-Receptor Blocker. Hypertension 50: 89-95 [Abstract] [Full text]  
  • Authors/Task Force Members:, , Mancia, G., De Backer, G., Dominiczak, A., Cifkova, R., Fagard, R., Germano, G., Grassi, G., Heagerty, A. M., Kjeldsen, S. E., Laurent, S., Narkiewicz, K., Ruilope, L., Rynkiewicz, A., Schmieder, R. E., Struijker Boudier, H. A.J., Zanchetti, A., ESC Committee for Practice Guidelines (CPG):, , Vahanian, A., Camm, J., De Caterina, R., Dean, V., Dickstein, K., Filippatos, G., Funck-Brentano, C., Hellemans, I., Kristensen, S. D., McGregor, K., Sechtem, U., Silber, S., Tendera, M., Widimsky, P., Zamorano, J. L., ESH Scientific Council:, , Kjeldsen, S. E., Erdine, S., Narkiewicz, K., Kiowski, W., Agabiti-Rosei, E., Ambrosioni, E., Cifkova, R., Dominiczak, A., Fagard, R., Heagerty, A. M., Laurent, S., Lindholm, L. H., Mancia, G., Manolis, A., Nilsson, P. M., Redon, J., Schmieder, R. E., Struijker-Boudier, H. A.J., Viigimaa, M., Document Reviewers:, , Filippatos, G., Adamopoulos, S., Agabiti-Rosei, E., Ambrosioni, E., Bertomeu, V., Clement, D., Erdine, S., Farsang, C., Gaita, D., Kiowski, W., Lip, G., Mallion, J.-M., Manolis, A. J., Nilsson, P. M., O'Brien, E., Ponikowski, P., Redon, J., Ruschitzka, F., Tamargo, J., van Zwieten, P., Viigimaa, M., Waeber, B., Williams, B., Zamorano, J. L. (2007). 2007 Guidelines for the Management of Arterial Hypertension: The Task Force for the Management of Arterial Hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC). Eur Heart J 0: ehm236v1-75 [Full text]  
  • Choi, J.-H., Xu, Q.-S., Park, S.-Y., Kim, J.-H., Hwang, S.-S., Lee, K.-H., Lee, H.-J., Hong, Y.-C. (2007). Seasonal variation of effect of air pollution on blood pressure. J. Epidemiol. Community Health 61: 314-318 [Abstract] [Full text]  
  • Chiolero, A., Bovet, P., Paradis, G., Paccaud, F. (2007). Has Blood Pressure Increased in Children in Response to the Obesity Epidemic?. Pediatrics 119: 544-553 [Abstract] [Full text]  
  • Chen, Y., Factor-Litvak, P., Howe, G. R., Graziano, J. H., Brandt-Rauf, P., Parvez, F., van Geen, A., Ahsan, H. (2007). Arsenic Exposure from Drinking Water, Dietary Intakes of B Vitamins and Folate, and Risk of High Blood Pressure in Bangladesh: A Population-based, Cross-sectional Study. Am J Epidemiol 165: 541-552 [Abstract] [Full text]  
  • Singhal, A., Cole, T. J., Fewtrell, M., Kennedy, K., Stephenson, T., Elias-Jones, A., Lucas, A. (2007). Promotion of Faster Weight Gain in Infants Born Small for Gestational Age: Is There an Adverse Effect on Later Blood Pressure?. Circulation 115: 213-220 [Abstract] [Full text]  
  • Chen, Y., Factor-Litvak, P., Howe, G. R, Parvez, F., Ahsan, H. (2006). Nutritional influence on risk of high blood pressure in Bangladesh: a population-based cross-sectional study.. Am. J. Clin. Nutr. 84: 1224-1232 [Abstract] [Full text]  
  • Stang, A., Moebus, S., Mohlenkamp, S., Dragano, N., Schmermund, A., Beck, E.-M., Siegrist, J., Erbel, R., Jockel, K.-H., on behalf of the Heinz Nixdorf Recall Study Invest, (2006). Algorithms for Converting Random-Zero to Automated Oscillometric Blood Pressure Values, and Vice Versa. Am J Epidemiol 164: 85-94 [Abstract] [Full text]  
  • Schell, K., Lyons, D., Bradley, E., Bucher, L., Seckel, M., Wakai, S., Carson, E., Waterhouse, J., Chichester, M., Bartell, D., Foraker, T., Simpson, E. K. (2006). Clinical Comparison of Automatic, Noninvasive Measurements of Blood Pressure in the Forearm and Upper Arm With the Patient Supine or With the Head of the Bed Raised 45{degrees}: A Follow-Up Study. Am J Crit Care 15: 196-205 [Abstract] [Full text]  
  • Findlay, J. Y., Gali, B., Keegan, M. T., Burkle, C. M., Plevak, D. J. (2006). Vasotrac(R) arterial blood pressure and direct arterial blood pressure monitoring during liver transplantation.. Anesth. Analg. 102: 690-693 [Abstract] [Full text]  
  • Gravlee, C. C., Dressler, W. W., Bernard, H. R. (2005). Skin Color, Social Classification, and Blood Pressure in Southeastern Puerto Rico. Am. J. Public Health 95: 2191-2197 [Abstract] [Full text]  
  • Olsen, M. H., Hansen, T. W., Christensen, M. K., Gustafsson, F., Rasmussen, S., Wachtell, K., Borch-Johnsen, K., Ibsen, H., Jorgensen, T., Hildebrandt, P. (2005). N-Terminal Pro Brain Natriuretic Peptide Is Inversely Related to Metabolic Cardiovascular Risk Factors and the Metabolic Syndrome. Hypertension 46: 660-666 [Abstract] [Full text]  
  • Veldhuijzen van Zanten, J J C S, Ring, C, Carroll, D, Kitas, G D (2005). Increased C reactive protein in response to acute stress in patients with rheumatoid arthritis. Ann Rheum Dis 64: 1299-1304 [Abstract] [Full text]  
  • Kaplan, G. A., Siefert, K., Ranjit, N., Raghunathan, T. E., Young, E. A., Tran, D., Danziger, S., Hudson, S., Lynch, J. W., Tolman, R. (2005). The Health of Poor Women Under Welfare Reform. Am. J. Public Health 95: 1252-1258 [Abstract] [Full text]  
  • Verbeke, F., Segers, P., Heireman, S., Vanholder, R., Verdonck, P., Van Bortel, L. M. (2005). Noninvasive Assessment of Local Pulse Pressure: Importance of Brachial-to-Radial Pressure Amplification. Hypertension 46: 244-248 [Abstract] [Full text]  
  • Phillips, A. C., Carroll, D., Ring, C., Sweeting, H., West, P. (2005). Life Events and Acute Cardiovascular Reactions to Mental Stress: A Cohort Study. Psychosom. Med. 67: 384-392 [Abstract] [Full text]  
  • Hansen, T. W., Jeppesen, J., Rasmussen, S., Ibsen, H., Torp-Pedersen, C. (2005). Ambulatory Blood Pressure and Mortality: A Population-Based Study. Hypertension 45: 499-504 [Abstract] [Full text]  
  • Wong, W. C W, Shiu, I. K L, Hwong, T. M T, Dickinson, J. A (2005). Reliability of automated blood pressure devices used by hypertensive patients. JRSM 98: 111-113 [Abstract] [Full text]  
  • Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., Roccella, E. J. (2005). Recommendations for Blood Pressure Measurement in Humans and Experimental Animals: Part 1: Blood Pressure Measurement in Humans: A Statement for Professionals From the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Circulation 111: 697-716 [Abstract] [Full text]  
  • Celis, H., Hond, E. D., Staessen, J. A. (2005). Self-Measurement of Blood Pressure at Home in the Management of Hypertension. Clin Med Res 3: 19-26 [Abstract] [Full text]  
  • Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., Roccella, E. J. (2005). Recommendations for Blood Pressure Measurement in Humans and Experimental Animals: Part 1: Blood Pressure Measurement in Humans: A Statement for Professionals From the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension 45: 142-161 [Abstract] [Full text]  
  • Taegtmeyer, A. B., Crook, A. M., Barton, P. J.R., Banner, N. R. (2004). Reduced incidence of hypertension after heterotopic cardiac transplantation compared with orthotopic cardiac transplantation: Evidence that excision of the native heart contributes to post-transplant hypertension. J Am Coll Cardiol 44: 1254-1260 [Abstract] [Full text]  
  • Tholl, U., Forstner, K., Anlauf, M. (2004). Measuring blood pressure: pitfalls and recommendations. Nephrol Dial Transplant 19: 766-770 [Full text]  
  • Staessen, J. A., Den Hond, E., Celis, H., Fagard, R., Keary, L., Vandenhoven, G., O'Brien, E. T. (2004). Antihypertensive Treatment Based on Blood Pressure Measurement at Home or in the Physician's Office: A Randomized Controlled Trial. JAMA 291: 955-964 [Abstract] [Full text]  
  • White, J. R. Jr., Schick, J. (2004). Home Blood Pressure Monitoring and Diabetes. Clin. Diabetes 22: 28-31 [Full text]  
  • Chang, J. J., Rabinowitz, D., Shea, S. (2003). Sources of Variability in Blood Pressure Measurement using the Dinamap PRO 100 Automated Oscillometric Device. Am J Epidemiol 158: 1218-1226 [Abstract] [Full text]  
  • Carroll, D., Ring, C., Hunt, K., Ford, G., Macintyre, S. (2003). Blood Pressure Reactions to Stress and the Prediction of Future Blood Pressure: Effects of Sex, Age, and Socioeconomic Position. Psychosom. Med. 65: 1058-1064 [Abstract] [Full text]  
  • Wolf-Maier, K., Cooper, R. S., Banegas, J. R., Giampaoli, S., Hense, H.-W., Joffres, M., Kastarinen, M., Poulter, N., Primatesta, P., Rodriguez-Artalejo, F., Stegmayr, B., Thamm, M., Tuomilehto, J., Vanuzzo, D., Vescio, F. (2003). Hypertension Prevalence and Blood Pressure Levels in 6 European Countries, Canada, and the United States. JAMA 289: 2363-2369 [Abstract] [Full text]  
  • Jones, D. W., Appel, L. J., Sheps, S. G., Roccella, E. J., Lenfant, C. (2003). Measuring Blood Pressure Accurately: New and Persistent Challenges. JAMA 289: 1027-1030 [Full text]  
  • Bur, A., Herkner, H., Vlcek, M., Woisetschlager, C., Derhaschnig, U., Hirschl, M. M. (2002). Classification of Blood Pressure Levels by Ambulatory Blood Pressure in Hypertension. Hypertension 40: 817-822 [Abstract] [Full text]  
  • Cockcroft, J. R., Wilkinson, I. B. (2002). reliability of transfer function in determining central pulse pressure and augmentation index: Reply. J Am Coll Cardiol 40: 1196-1197 [Full text]  
  • Campbell, N. R.C., Drouin, D., Feldman, R. D. (2002). The 2001 Canadian hypertension recommendations: take-home messages. CMAJ 167: 661-668 [Full text]  
  • Lewis, J. E., Boyle, E., Magharious, L., Myers, M. G. (2002). Evaluation of a community-based automated blood pressure measuring device. CMAJ 166: 1145-1148 [Abstract] [Full text]  
  • Quinn, M., Michell, A R, O'Brien, E., Lip, G. Y H, Beevers, D G., Marshall, T., Rouse, A. (2001). Blood pressure measurement. BMJ 323: 805-805 [Full text]  
  • O'Brien, E., Jones, D. W., Frohlich, E. D., Grim, C. M., Grim, C. E., Taubert, K. A. (2001). Response to the Advisory Statement From the Council for High Blood Pressure Research of the American Heart Association Advocating Retention of Mercury Sphygmomanometers Response. Hypertension 38 : e19-e20 [Full text]  
  • Thomas, M., Radford, T., Dasgupta, I. (2001). Unvalidated blood pressure devices with small cuffs are being used in hospitals. BMJ 323: 398-398 [Full text]  
  • Gruskin, A. B. (2001). Blood Pressure: Dinamap Versus Sphygmomanometer. AAP Grand Rounds 5: 47-48 [Full text]  
  • Law, C.M., Shiell, A.W., Newsome, C.A., Syddall, H.E., Shinebourne, E.A., Fayers, P.M., Martyn, C.N., de Swiet, M. (2002). Fetal, Infant, and Childhood Growth and Adult Blood Pressure: A Longitudinal Study From Birth to 22 Years of Age. Circulation 105: 1088-1092 [Abstract] [Full text]  

Rapid Responses:

Read all Rapid Responses

Brief Inquiry about Omron new model HEM-741.
Harold Sheridan Adams
bmj.com, 3 Mar 2001 [Full text]
Unvalidated BP devices with small cuffs are used on hospital wards
Mark Thomas
bmj.com, 28 Mar 2001 [Full text]
Aneroid gauges are accurate in USA
Phillip J Colquitt
bmj.com, 6 Apr 2001 [Full text]
Blood Pressure Measurement Should be Taken With The Arm Straight
C P R Williams
bmj.com, 13 Apr 2001 [Full text]
Aneroidosis finds it’s way into the reluctant UK
Phillip J. Colquitt
bmj.com, 5 Jul 2003 [Full text]